Actuator LN Valve Creation

From Aeon HACS Wiki
Revision as of 11:39, 26 July 2024 by Oscar (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
    3D Model    About this item    Product info

Instructions

Actuator LN valve exploded
Actuator LN valve exploded
  1. Open the serial port terminal.
    Ensure the serial port terminal is open on your computer.
  2. Connect the Test Actuator Controller.
    Plug the test actuator controller into the computer.
  3. Connect the servo.
    Attach the CPW servo to the controller.
  4. Command the servo.
    Enter the command 'g1000' in the terminal to initialize the servo.
  5. Disconnect the servo.
    Safely disconnect the servo from the controller.
  6. Install the LN valve stem
    Attach the LN valve stem onto the servo drive spline, ensuring the orientation mark is facing the servo label.
  7. Install the LN valve stem collar
    Place the LN valve stem collar onto the valve stem with the chamfered ID first.
  8. Secure the collar
    using the collar pusher tool (TL-LNVSCP), press the collar down into contact with the servo gear cover.

Notes:

  • Do not assemble the screws and servo T-nut onto the valve at this stage. These will be assembled during the LN manifold assembly process.